25. Other Input Related Spurious Responses, HP 8593EM

Band 4

24.On the EMC analyzer, press 4FREQUENCY5, 21, 4GHz5.

25.Set the synthesized sweeper CW to 21 GHz.

26.Enter the power sensor 21 GHz CAL Factor into the measuring receiver.

27.Press the following EMC analyzer keys:

4MKR !5 MKR ! HIGH

4AMPLITUDE5 More 1 of 3 PRESEL PEAK

Wait for the CAL: PEAKING message to disappear, then press:

4MKR5 More 1 of 3 MARKER ALL OFF

28.Repeat steps 5 through 8 for the synthesized sweeper CW frequencies listed in Table 2-43 for Band 4.

Band 4 for Option 026 or 027

Perform this section only if you EMC analyzer is equipped with Option 026 or Option 027.

29.On the EMC analyzer, press 4FREQUENCY5, 24, 4GHz5.

30.Set the synthesized sweeper CW to 24 GHz.

31.Enter the power sensor 24 GHz CAL Factor into the measuring receiver.

32.Press the following EMC analyzer keys:

4MKR !5 MKR ! HIGH

4AMPLITUDE5 More 1 of 3 PRESEL PEAK

Wait for the CAL: PEAKING message to disappear, then press:

4MKR5 More 1 of 3 MARKER ALL OFF

33.Repeat steps 5 through 8 for the synthesized sweeper CW frequencies listed in Table 2-43 for Band 4 for Option 026 or Option 027.

2-162 Performance Veri￿cation Tests