25. Other Input Related Spurious Responses, HP 8593EM
Band 4
24.On the EMC analyzer, press 4FREQUENCY5, 21, 4GHz5.
25.Set the synthesized sweeper CW to 21 GHz.
26.Enter the power sensor 21 GHz CAL Factor into the measuring receiver.
27.Press the following EMC analyzer keys:
4MKR !5 MKR ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK
Wait for the CAL: PEAKING message to disappear, then press:
4MKR5 More 1 of 3 MARKER ALL OFF
28.Repeat steps 5 through 8 for the synthesized sweeper CW frequencies listed in Table
Band 4 for Option 026 or 027
Perform this section only if you EMC analyzer is equipped with Option 026 or Option 027.
29.On the EMC analyzer, press 4FREQUENCY5, 24, 4GHz5.
30.Set the synthesized sweeper CW to 24 GHz.
31.Enter the power sensor 24 GHz CAL Factor into the measuring receiver.
32.Press the following EMC analyzer keys:
4MKR !5 MKR ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK
Wait for the CAL: PEAKING message to disappear, then press:
4MKR5 More 1 of 3 MARKER ALL OFF
33.Repeat steps 5 through 8 for the synthesized sweeper CW frequencies listed in Table