22. Frequency Response, HP 8595EM

5.On the EMC analyzer, press 4MKR !5 MARKER ! HIGH , 4MKR5,

More 1 of 3 , and MK TRACK ON OFF (ON).

6.Adjust the synthesized sweeper power level for a MKR-TRK amplitude reading of 93 dB￿V 60.1 dB.

7.Press RATIO on the measuring receiver.

Frequency Response , Band 0, ￿50 MHz

8.Set the synthesized sweeper CW FREQUENCY to 50 MHz.

9.Set the EMC analyzer CENTER FREQUENCY to 50 MHz.

10.Adjust the synthesized sweeper power level for an EMC analyzer MKR-TRK amplitude reading of 93 dB￿V 60.1 dB.

11.Record the power ratio displayed on the measuring receiver below, then record the negative of this value in column 2 of Table 2-35 as the Measuring Receiver Reading at 50 MHz.

Measuring Receiver Reading at 50 MHz

 

dB

12.Set the synthesized sweeper CW FREQUENCY to 100 MHz.

13.Set the EMC analyzer CENTER FREQUENCY to 100 MHz.

14.Adjust the synthesized sweeper power level for an EMC analyzer MKR-TRK amplitude reading of 93 dB￿V 60.1 dB.

15.Record the negative of the power ratio displayed on the measuring receiver in Table 2-35 as the measuring receiver Reading.

16.On the synthesized sweeper, press 4CW5 , and 4 * 5 (step up) key and on the EMC analyzer, press 4FREQUENCY5, 4 * 5 (step up) key to step through the remaining frequencies listed in Table 2-35.

17.At each new frequency repeat steps 13 through 15, entering the power sensor Cal Factor into the measuring receiver as indicated in Table 2-35.

Performance Veri￿cation Tests 2-133