17.Resolution (IF) Bandwidth Accuracy, HP 8590
EMC Analyzer |
| Column 1 |
| Column 2 | TR Entry | ||
RES BW |
| Synthesizer |
| Synthesizer |
| ||
|
| Lower Frequency |
| Upper Frequency |
| ||
|
|
|
|
|
|
|
|
3 MHz |
|
|
|
|
|
|
|
|
|
|
|
|
| 1 | |
300 kHz |
|
|
|
|
|
| 2 |
100 kHz |
|
|
|
|
|
| 3 |
30 kHz |
|
|
|
|
|
| 4 |
10 kHz |
|
|
|
|
|
| 5 |
3 kHz |
|
|
|
|
|
| 6 |
1 kHz |
|
|
|
|
|
| 7 |
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
6 dB EMI Bandwidths
17.Set the synthesizer/level generator AMPTD INCR to 6 dB.
18.On the EMC analyzer, press the following keys:
4BW5 9 kHz EMI BW
4MKR5 MARKER NORMAL
19.On the synthesizer/level generator, press FREQUENCY. Adjust the frequency for a maximum marker reading.
20.On the synthesizer/level generator, press AMPLITUDE and INCR 4+5
21.Press MARKER 1 on the EMC analyzer.
22.On the synthesizer/level generator, press INCR 4*5
23.On the synthesizer/level generator, press FREQUENCY. Lower the frequency of the synthesizer/level generator by adjusting the knob until the
24.Record the synthesizer/level generator frequency readout in column 1 of Table