16.Absolute Amplitude Calibration and Resolution (IF) Bandwidth Switching Uncertainties,HP 8590
4SPAN5 50 4kHz5
4BW5 3 4kHz5
4MKR !5 MARKER ! HIGH
More 1 of 3 More 2 of 3 MARKER 1
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
10.Set the bandwidth and span according to Table
11.Press 4MKR !5, MARKER ! HIGH , then record the MKR 1 TRK amplitude reading in the performance verication test record as indicated in Table
The amplitude reading should be within the limits shown.
12.Repeat steps 10 through 11 for each of the remaining bandwidth and span settings listed in Table
Table
Resolution (IF) Bandwidth Switching Uncertainty for
Narrow Bandwidths
EMC Analyzer | MKR 1 TRK Amplitude Reading |
|
|
IF BW Setting SPAN Setting Min. (dB) TR Entry Max. (dB)
3 kHz | 50 kHz | 0 (Ref) | 0 (Ref) | 0 (Ref) | |
300 | Hz | 1 kHz | 00.6 | 11 | +0.6 |
200 Hz | 1 kHz | 00.6 | 12 | +0.6 | |
100 Hz | 1 kHz | 00.6 | 13 | +0.6 | |
30 Hz | 1 kHz | 00.6 | 14 | +0.6 | |
|
|
|
|
|
|
Note that it is normal for the 200 Hz bandwidth shape to have a dip in the center of the response.