53. Tracking Generator Level Flatness,
HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM Option 010
14.Set the DVM to measure AC Volts. Press the following DVM keys so that it reads out in dBm:
50 4STORE5 4
4MATH5 4
15.Set the EMC analyzer center frequency to 9 kHz and press 4SGL SWP5. Record the DVM readout in column 2 of Table
16.Repeat step 15 for all center frequencies listed in Table
Table
Tracking Generator Level Flatness Worksheet, <100 kHz
Center | DVM | Corrected |
Frequency | Readout | Level Flatness |
| (dBm) | (dBm) |
|
|
|
9kHz
20 kHz
40 kHz
60 kHz
80 kHz
100 kHz
17.Subtract the 100 kHz Level Flatness readout in Table
DVM Oset |
| dB |
|
18.For example, if the Level Flatness reading from Table
(DVM) 0 (Power Meter) = DVM Oset
19.Add the DVM Oset from Step 16 to each of the DVM Readouts in Table
For example, if the DVM Readout from Table
(DVM) + (DVM Oset) = Corrected Readout