10. Residual FM, HP 8591EM
Part 1: Residual FM
Determining the IF Filter Slope
1.Connect the equipment as shown in Figure
2.Set the signal generator controls as follows:
FREQUENCY ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: : 500 MHz
CW OUTPUT :: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: 010 dBm
3.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 500 4MHz5
4SPAN5 1 4MHz5
4AMPLITUDE5 98 4dBV5
SCALE LOG LIN (LOG) 1 4dB5
4BW5 1 4kHz5
4MKR !5 MKR ! HIGH
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4SPAN5 10 4kHz5
4.Wait for the AUTO ZOOM message to disappear, then press the following EMC analyzer keys:
4MKR !5 More 1 of 3 MARKER !REF LVL
4MKR5 MARKER 1 ON OFF (OFF)
4SGL SWP5
4MKR !5 MKR ! HIGH More 1 of 3 More 2 of 3 MARKER 1
If you have diculty achieving the 60.1 dB setting, then make the following EMC analyzer settings:
4SPAN5 5 4kHz5
4BW5 AVG BW AUTO MAN (MAN) 30 4Hz5
5.Rotate the EMC analyzer knob counterclockwise until the
amplitude reads 01 dB 60.1 dB. Press 4MKR5, MARKER 1 . Rotate the knob counterclockwise until the
Performance Verication Tests