If your instrument is equipped with a Precision Frequency Reference, perform \10 MHz Precision Frequency Reference Output Accuracy," instead.
This performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
The settability is measured by changing the setting of the
The related adjustment for this performance verication test is the \10 MHz Reference, HP 8590
Equipment Required
Microwave frequency counter
Frequency standard
Cable, BNC, 122 cm (48 in) (2 required)
Figure
Procedure
The test results will be invalid if REF UNLK is displayed at any time during this test. REF UNLK will be displayed if the internal reference oscillator is unlocked from the 10 MHz reference. A REF UNLK might occur if there is a hardware failure or if the jumper between 10 MHz REF OUTPUT and EXT REF IN on the rear panel is removed.