53. Tracking Generator Level Flatness, HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM Option 010
6.Disconnect the Type N cable from the RF OUT 50 and connect the 100 kHz to 2.9 GHz power sensor to the RF OUT 50 .
7. On the EMC analyzer, press SRC PWR ON OFF (ON), 87 4dBV5, 4SGL SWP5.
8.Press RATIO on the measuring receiver. The measuring receiver readout is now in power levels relative to the power level at 300 MHz.
9.Set the EMC analyzer center frequency to 100 kHz. Press 4SGL SWP5.
10.Enter the appropriate power sensor Cal Factor into the measuring receiver as indicated in Table
11.Record the power level displayed on the measuring receiver as the Level Flatness in Table
12.Repeat steps 9 through 11 to measure the atness at each center frequency setting listed in Table
Performance Verication Tests