HP 8593EM Performance Test Record

HP 8593EM Performance Test Record (page 10 of 13)

Hewlett-P ackard Company

 

 

 

 

 

 

 

 

 

 

 

 

 

Model HP 8593EM

 

 

Report No .

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Serial No .

 

 

 

Date

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Test Description

 

 

Results Measured

 

 

Measurement

 

 

 

 

 

 

Min.

 

 

 

(TR Entry)

Max.

 

Uncertainty

20. Frequency Response

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Band 0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Max Positive Response

 

 

(1)

 

 

 

 

 

+1.5 dB

+0.32/ 00.33 dB

 

 

 

 

 

 

 

 

Max Negative Response

01.5 dB

(2)

 

 

 

 

 

 

 

+0.32/ 00.33 dB

 

Peak-to-P eak Response

 

 

(3)

 

 

 

 

 

2.0

dB

+0.32/ 00.33 dB

 

 

 

 

 

 

 

 

Band 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Max Positive Response

 

 

(4)

 

 

 

 

 

+2.0

dB

+0.40/ 00.42 dB

 

 

 

 

 

 

 

 

Max Negative Response

02.0 dB

(5)

 

 

 

 

 

 

 

+0.40/ 00.42 dB

 

Peak-to-P eak Response

 

 

(6)

 

 

 

 

 

3.0

dB

+0.40/ 00.42 dB

 

 

 

 

 

 

 

 

Band 2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Max Positive Response

 

 

(7)

 

 

 

 

 

+2.5

dB

+0.42/ 00.43 dB

 

 

 

 

 

 

 

 

Max Negative Response

02.5 dB

(8)

 

 

 

 

 

 

 

+0.42/ 00.43 dB

 

Peak-to-P eak Response

 

 

(9)

 

 

 

 

 

4.0 dB

+0.42/ 00.43 dB

 

 

 

 

 

 

 

 

Band 3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Max Positive Response

 

 

(10)

 

 

 

 

 

+3.0

dB

+0.52/ 00.55 dB

 

 

 

 

 

 

 

 

Max Negative Response

03.0 dB

(11)

 

 

 

 

 

 

 

+0.52/ 00.55 dB

 

 

 

 

 

 

 

 

Peak-to-P eak Response

 

 

(12)

 

 

 

 

 

4.0

dB

+0.52/ 00.55 dB

Band 4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Max Positive Response

 

 

(13)

 

 

 

 

 

+3.0 dB

+0.54/ 00.57 dB

Max Negative Response

03.0 dB

(14)

 

 

 

 

 

 

 

+0.54/ 00.57 dB

 

 

 

 

 

 

 

 

Peak-to-P eak Response

 

 

(15)

 

 

 

 

 

4.0 dB

+0.54/ 00.57 dB

Band 4 for Option 026 or 027

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Max Positive Response

 

 

(13)

 

 

 

 

 

+5.0 dB

+0.54/ 00.57 dB

Max Negative Response

05.0 dB

(14)

 

 

 

 

 

 

 

+0.54/ 00.57 dB

 

 

 

 

 

 

 

 

Peak-to-P eak Response

 

 

(15)

 

 

 

 

 

4.0

dB

+0.54/ 00.57 dB

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

25.Other Input Related Spurious Responses

50 kHz to 2.9 GHz

55 dBc

(1)

 

 

 

 

 

+1.12/ 01.21 dB

 

 

 

 

 

 

￿18 GHz

55 dBc

(2)

 

 

 

 

 

+1.13/ 01.22 dB

 

￿22 GHz

50 dBc

(3)

 

 

 

 

 

+1.15/ 01.25 dB

 

 

 

 

 

 

Option 026 or 027 only:

 

 

 

 

 

 

 

 

￿26.5 GHz

50 dBc

(3)

 

 

 

 

 

+1.15/ 01.25 dB

3-26

Performance

Test Records

 

 

 

 

 

 

 

30. Spurious Response

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Second Harmonic

Distortion

 

 

 

 

 

 

 

 

Applied Frequency

 

 

 

 

 

 

 

 

40 MHz

 

(1)

 

 

050 dBc

+1.86/ 02.27 dB

 

 

 

2.8 GHz

 

(3)

 

 

(2)

 

 

+2.24/ 02.72 dB

 

 

 

Third Order Intermodulation

 

 

 

 

(Step 23c)

 

 

Distortion

 

 

 

 

 

 

 

 

Frequency