30. Spurious Response, HP 8593EM
55. Set the EMC analyzer by pressing the following keys:
4BW5 IF BW AUTO MAN (MAN) 1 4kHz5
AVG BW AUTO MAN (MAN) 100 4Hz5
56.Press 4MKR !5 MARKER ! HIGH , More 1 of 3 More 2 of 3 MARKER 1 then set the DISPLAY LINE to a value 54 dB below the current reference level setting. The
57.If the distortion products can be seen, proceed as follows:
a.On the EMC analyzer, press 4MKR !5, MARKER ! HIGH , More 1 of 3 ,
More 2 of 3 , and MARKER 1 .
b.Press More 3 of 3 then repeatedly press NEXT PEAK until the active marker is on the highest distortion product.
c.Record the MKR 1 amplitude reading as TR Entry 2 of the performance verication test record. The MKR 1 reading should be less than 054 dBc.
58.If the distortion products cannot be seen, proceed as follows:
a.On both the synthesized sweeper and the synthesized level generator, increase the POWER LEVEL by 5 dB. Distortion products should now be visible at this higher power level.
b.On the EMC analyzer, press 4MKR !5, MARKER ! HIGH , More 1 of 3 ,
More 2 of 3 , and MARKER 1 .
c.Press More 3 of 3 then repeatedly press NEXT PEAK until the active marker is on the highest distortion products.
d.On both the synthesized sweeper and the synthesizer level generator, reduce the power level by 5 dB and wait for the completion of a new sweep.
e.Record the MKR 1 amplitude reading as TR Entry 2 of the performance verication test record. The MKR 1 reading should be less than 054 dBc.