51. Power Sweep Range, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM
11.Press 4AUX/USER5, Track Gen , PWR SWP ON OFF (OFF) to set power sweep
o. Press SRC PWR ON OFF (ON) and adjust the SRC POWER level until the 1MKR reads 01 dB 60.1 dB.
Be sure to wait for the completion of a new sweep after each adjustment of the SRC POWER level.
12.Record the power level displayed on the measuring receiver as TR Entry 2 of the performance verication test record.
13.Subtract Start Power Level (TR Entry 1) from the Stop Power Level
(TR Entry 2) and record as the Power Sweep Range in the performance verication test record as TR Entry 3.
Power Sweep Range = Stop Power Level 0 Start Power Level