51. Power Sweep Range, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM

11.Press 4AUX/USER5, Track Gen , PWR SWP ON OFF (OFF) to set power sweep

o￿. Press SRC PWR ON OFF (ON) and adjust the SRC POWER level until the 1MKR reads 01 dB 60.1 dB.

Be sure to wait for the completion of a new sweep after each adjustment of the SRC POWER level.

12.Record the power level displayed on the measuring receiver as TR Entry 2 of the performance veri￿cation test record.

13.Subtract Start Power Level (TR Entry 1) from the Stop Power Level

(TR Entry 2) and record as the Power Sweep Range in the performance veri￿cation test record as TR Entry 3.

Power Sweep Range = Stop Power Level 0 Start Power Level

2-334 Performance Veri￿cation Tests