35. Gain Compression, HP 8593EM

Option 026 only: Connect the directional coupler to the EMC analyzer directly.

9. On the EMC analyzer, press the following keys:

4MKR !5 MKR ! HIGH

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

4SPAN5 10 4MHz5

Wait for the AUTO ZOOM routine to ￿nish.

10.On synthesized sweeper #2, adjust the power level to place the signal 1 dB below the EMC analyzer reference level.

11.On the EMC analyzer, press 4MKR !5 MKR ! HIGH More 1 of 3

More 2 of 3 MARKER 1 .

12.On synthesized sweeper #1, set RF to ON.

13.On the EMC analyzer, press 4MKR !5 MKR ! HIGH , then NEXT PEAK .

The active marker should be on the lower amplitude signal and not on the signal that is o￿ the top of the screen. If it is not on the lower amplitude signal, reposition the marker to this peak using the EMC analyzer knob.

14.Read the MKR 1 amplitude and record in the performance veri￿cation test record as TR Entry 1. The absolute value of this amplitude should be less than 0.5 dB.

Gain Compression, >2.9 GHz

15.Disconnect the directional coupler from the EMC analyzer input, then connect the directional coupler to the power sensor.

16.Set the EMC analyzer by pressing the following key:

4FREQUENCY5 4.0 4GHz5

4SPAN5 20 4MHz5

4MKR5 More 1 of 3 MARKER ALL OFF

2-244 Performance Veri￿cation Tests