8.Frequency Span Readout Accuracy, HP 8591EM Table
EMC Analyzer | Synthesizer/Level | Synthesized Sweeper |
| |||
Span Setting | Generator | Frequency |
|
|
|
|
| Frequency |
|
|
|
|
|
|
|
|
|
|
| |
| MHz | MHz | Minimum | TR Entry | Maximum | |
|
|
|
|
|
| |
10.10 MHz | 66.000 | 74.000 | 7.70 MHz | 2 | 8.30 MHz | |
10.00 MHz | 66.000 | 74.000 | 7.80 MHz | 3 | 8.20 MHz | |
100.00 kHz | 69.960 | 70.040 | 78.00 kHz | 4 | 82.00 kHz | |
99.00 kHz | 69.960 | 70.040 | 78.00 kHz | 5 | 82.06 kHz | |
10.00 kHz | 69.996 | 70.004 | 7.80 | kHz | 6 | 8.20 kHz |
|
|
|
|
|
|
|
1.00 kHz | 69.9996 | 70.0004 | 0.78 | kHz | 7 | 0.82 kHz |
300.00 Hz 1 | 69.99988 | 70.00012 | 225.00 | Hz | 255. 00 Hz |
1 This is not an EMC analyzer specication; however, the 300 Hz span is tested to 65 % to keep the narrow bandwidth accuracy and residual FM measurement uncertainty at a minimum. If the 300 Hz span accuracy is >5% the additional measurement uncertainty may need to be included for the bandwidth accuracy and residual FM measurement uncertainties .
Performance Verication Tests