51. Power Sweep Range, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM
Procedure
1.Connect the equipment as shown in Figure
2.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 Band Lock
The HP 8594EM does not need to be band locked.
4FREQUENCY5 300 4MHz5
4SPAN5 0 4Hz5
4BW5 IF BW AUTO MAN (MAN) 30 4kHz5
4MKR5
4AUX/USER5 Track Gen SRC PWR ON OFF (ON) 102 4dBV5
3.On the EMC analyzer, press TRACKING PEAK , then wait for the PEAKING! message to disappear.
4.Zero and calibrate the
Connect the power sensor to the power splitter. See Figure
5.On the EMC analyzer, press the following keys:
SRC PWR ON OFF (ON) 97 4dBV5
SRC ATN MAN AUTO (MAN) 0 4dB5
PWR SWP ON OFF (ON) 10 4dB5
4AMPLITUDE5 SCALE LOG LIN (LOG) 2 4dB5
Press REF LVL on the EMC analyzer, then adjust the reference level until the peak of the displayed ramp (along the
6.Press 4MKR5, MARKER NORMAL . Use the knob to place the marker at the
MARKER 1 .
7.Press 4AUX/USER5, Track Gen , PWR SWP ON OFF (OFF) to set power sweep o. The 1MKR should read 0 dB 60.1 dB. If it does not, press