4. Frequency Readout and Marker Count Accuracy, HP 8591EM and HP 8594EM
Part 2: Marker Count A ccuracy
Perform \Part 1: Frequency Readout Accuracy" before performing this procedure.
1.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer to measure the marker count accuracy by pressing the following keys:
4FREQUENCY5 1.5 4GHz5
4SPAN5 20 4MHz5
4BW5 IF BW AUTO MAN (MAN) 300 4kHz5
4MKR5 More 1 of 3 MK COUNT ON OFF (ON)
More 2 of 3 CNT RES AUTO MAN (MAN) 100 4Hz5
2.Press 4MKR !5 MARKER ! HIGH , then wait for a count to be taken (it may take several seconds).
3.Record the CNTR frequency reading as TR Entry 5 of the performance verication test record. The reading should be within the limits of 1.4999989 GHz and 1.5000011 GHz.
4.Change the EMC analyzer settings by pressing the following keys:
4SPAN5 1 4MHz5
4MKR5 More 1 of 3 More 2 of 3 CNT RES AUTO MAN (MAN) 10 4Hz5
5.Press 4MKR !5 MARKER ! HIGH , then wait for a count to be taken (it may take several seconds).
6.Record the CNTR frequency reading as TR Entry 6 of the performance verication test record. The reading should be within the limits of 1.49999989 GHz and 1.50000011 GHz.
7.Set the EMC analyzer by pressing the following keys:
4BW5 IF BW AUTO MAN (MAN) 300 4Hz5
4SPAN5 20 4kHz5
8.Press 4MKR !5 MARKER ! HIGH on the EMC analyzer.
9.Record the MKR frequency reading as TR Entry 7 of the performance verication test record. The reading should be within the limits of 1.49999989 GHz and 1.50000011 GHz.
Performance Verication Tests