40. Displayed Average Noise Level, HP 8593EM
4MKR !5 MARKER ! HIGH
4AMPLITUDE5 More 1 of 3 More 2 of 3 REF LVL OFFSET
Subtract the MKR amplitude reading from 87 dBV and enter the result as the REF LVL OFFSET. For example, if the marker reads 86.79 dBV, enter 00.21 dB (87 dBV 0 86.79 dBV = +0.21 dB).
REF LVL OFFSET |
| dB |
5.Disconnect the cable from the INPUT 50 connector of the EMC analyzer. Connect the 50 termination to the EMC analyzer INPUT 50 connector.
Performance Verication Tests