21. Frequency Response, HP 8594EM
6.Adjust the synthesized sweeper power level for a
7.Set the power sensor cal factor for the measuring receiver, then press
RATIO.
8.Set the synthesized sweeper CW to 50 MHz.
9.Press 4FREQUENCY5, 50 4MHz5 on the EMC analyzer.
10.Adjust the synthesized sweeper power level for an EMC analyzer
11.Set the power sensor cal factor for the measuring receiver, then record the power ratio displayed on the measuring receiver below. Record the negative of the power ratio in Table
Measuring Receiver Reading at 50 MHz |
| dB |
12.Set the synthesized sweeper CW to 100 MHz.
13.Press 4FREQUENCY5, 100 4MHz5 on the EMC analyzer.
14.Adjust the synthesized sweeper power level for an EMC analyzer
15.Set the power sensor cal factor for the measuring receiver, then record the negative of the power ratio displayed on the measuring receiver in Table
16.On the synthesized sweeper, press CW, and 4 * 5 (step up) key.
Performance Verication Tests