30. Spurious Response, HP 8593EM
c.Record the more positive of the values recorded in steps a. and b. above as TR Entry 2 of the performance verication test record. For example, if the value in step a. is 059 dBc and the value in step b. is 061 dBc, record 059 dBc.
Figure
Second Harmonic Distortion Test Setup, >2.9 GHz, HP 8593EM
24.Connect the equipment as shown in Figure
25.Set the synthesized sweeper controls as follows:
CW :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: :: 2.8 GHz
POWER LEVEL ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: : 0 dBm
26. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 2.8 4GHz5
4MKR5 More 1 of 3 MARKER ALL OFF
4MKR !5 MARKER ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK
Wait for the peaking message to disappear.
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4SPAN5 100 4kHz5
Performance Verication Tests