33. Spurious Response, HP 8596EM
26. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 2.8 4GHz5
4MKR5 More 1 of 3 MARKER ALL OFF
4MKR !5 MARKER ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK Wait for the peaking message to disappear.
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4SPAN5 100 4kHz5
27.Adjust the synthesized sweeper power level for an EMC analyzer marker amplitude reading of 107 dBV 60.2 dB.
28.On the EMC analyzer, press the following keys:
4MKR5 More 1 of 3 MK TRACK ON OFF (OFF)
4MKR !5 MARKER ! HIGH More 1 of 3 More 2 of 3 MARKER 1
4FREQUENCY5 5.6 4GHz5
4SPAN5 10 4MHz5
29.Remove the lter and connect the synthesized sweeper output directly to the EMC analyzer INPUT 50 .
30.On the EMC analyzer, press the following keys:
4MKR !5 MARKER ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK
Wait for the peaking message to disappear.
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4SPAN5 100 4kHz5
31.Reinstall the lter between the synthesized sweeper output and the EMC analyzer INPUT 50 .
Performance Verication Tests