33. Spurious Response, HP 8596EM

26. Set the EMC analyzer by pressing the following keys:

4FREQUENCY5 2.8 4GHz5

4MKR5 More 1 of 3 MARKER ALL OFF

4MKR !5 MARKER ! HIGH

4AMPLITUDE5 More 1 of 3 PRESEL PEAK Wait for the peaking message to disappear.

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

4SPAN5 100 4kHz5

27.Adjust the synthesized sweeper power level for an EMC analyzer marker amplitude reading of 107 dB￿V 60.2 dB.

28.On the EMC analyzer, press the following keys:

4MKR5 More 1 of 3 MK TRACK ON OFF (OFF)

4MKR !5 MARKER ! HIGH More 1 of 3 More 2 of 3 MARKER 1

4FREQUENCY5 5.6 4GHz5

4SPAN5 10 4MHz5

29.Remove the ￿lter and connect the synthesized sweeper output directly to the EMC analyzer INPUT 50 ￿.

30.On the EMC analyzer, press the following keys:

4MKR !5 MARKER ! HIGH

4AMPLITUDE5 More 1 of 3 PRESEL PEAK

Wait for the peaking message to disappear.

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

4SPAN5 100 4kHz5

31.Reinstall the ￿lter between the synthesized sweeper output and the EMC analyzer INPUT 50 ￿.

Performance Veri￿cation Tests 2-229