62.CISPR Pulse Response, HP 8590
101.Enter the Measured value for the Band B 100 Hz Repetition Frequency as the Reference value for all the Repetition Frequencies listed for Band B.
102.Enter the Measured value for the Bands C and D 100 Hz Repetition Frequency as the Reference value for all the Repetition Frequencies listed for Bands C and D.
103.Calculate the error for each of the EMI bandwidths listed in Table
Measured CW Amplitude 0 Measured Amplitude for 25 Hz or 100 Hz = Error
104.Calculate the error for each of the frequencies listed in Table
Measured 0 Reference = Error
105.Calculate the error for each of the bands listed in Table
106.Record the calculated error values from Table
Table
EMI |
| Measured |
| Measured |
|
|
| Error | TR Entry | ||
Bandwidth |
| CW | Amplitude for |
|
|
|
| ||||
|
| Amplitude | 25 Hz or 100 | Hz |
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
|
|
200 Hz |
|
|
|
|
|
|
|
|
|
| 1 |
9 kHz |
|
|
|
|
|
|
|
|
|
| 2 |
120 kHz |
|
|
|
|
|
|
|
|
|
| 3 |
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
Performance Verication Tests