62.CISPR Pulse Response, HP 8590 EM-Series

101.Enter the Measured value for the Band B 100 Hz Repetition Frequency as the Reference value for all the Repetition Frequencies listed for Band B.

102.Enter the Measured value for the Bands C and D 100 Hz Repetition Frequency as the Reference value for all the Repetition Frequencies listed for Bands C and D.

103.Calculate the error for each of the EMI bandwidths listed in Table 2-81 and record in the Error column. Use the following formula to calculate the error:

Measured CW Amplitude 0 Measured Amplitude for 25 Hz or 100 Hz = Error

104.Calculate the error for each of the frequencies listed in Table 2-82 and record in the Error column. Use the following formula to calculate the error:

Measured 0 Reference = Error

105.Calculate the error for each of the bands listed in Table 2-83 and record in the Error column. Use the following formula to calculate the error:

Quasi-Peak Measurement 0 Average Measurement = Error

106.Record the calculated error values from Table 2-81, and Table 2-82, and Table 2-83 in the performance veri￿cation test record.

Table 2-81.

Quasi-Peak Detector Reference Accuracy Worksheet

EMI

 

Measured

 

Measured

 

 

 

Error

TR Entry

Bandwidth

 

CW

Amplitude for

 

 

 

 

 

 

Amplitude

25 Hz or 100

Hz

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

200 Hz

 

 

 

 

 

 

 

 

 

 

1

9 kHz

 

 

 

 

 

 

 

 

 

 

2

120 kHz

 

 

 

 

 

 

 

 

 

 

3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Performance Veri￿cation Tests 2-399