This performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
The tracking generator output is connected to the EMC analyzer input and the tracking is adjusted at 300 MHz for a maximum signal level. The tracking generator output is terminated and set for 106 dBV output power (maximum output power). The EMC analyzer input is also terminated. The noise level of the EMC analyzer is then measured at several frequencies.
There are no related adjustments for this performance verication test.
Equipment Required
Termination, 50 (two required)
Cable, Type N, 62 cm (24 in)
Cable, BNC, 23 cm (9 in)
Cable, Type N (m) to BNC (f)
Figure
Tracking Generator Feedthrough Test Setup,
HP 8593EM, HP 8595EM, and HP 8596EM