Table 1-6. Recommended Test Equipment

Equipment

 

 

Critical Speci￿cations for

Recommended

 

 

 

 

Equipment Substitution

Model

Digital V oltmeter

 

Input Resistance:

￿10 megohms

HP 3456A

 

 

 

Accuracy: 610 mV on 100 V range

 

Frequency Counter 1

 

Frequency: 10 MHz

 

HP 5334A/B

 

 

 

Resolution:

60.002 Hz

 

 

 

 

 

External Timebase

 

 

Frequency

Standard

 

Frequency:

10 MHz

 

HP 5061B

 

 

 

Timebase Accy (Aging): <1 21009/day

 

Measuring

Receiver

 

Compatible

with Power Sensors

HP 8902A

 

 

 

 

dB Relative Mode

 

 

 

 

 

 

Resolution:

0.01 dB

 

 

 

 

 

Reference

Accuracy: 61.2%

 

Microwave Frequency

 

Frequency Range:

9 MHz to 7 GHz

HP 5343A

 

Counter

 

 

Timebase Accy (Aging): <5 2 10010/day

 

Power Meter

 

Power Range: Calibrated

in dBm and

HP 436A

 

 

 

dB relative to reference

power 070 dBm

 

 

 

 

to +44 dBm, sensor dependent

 

Power Sensor

 

Frequency Range: 100 kHz to 1800 MHz

HP 8482A

 

 

 

 

Maximum SWR:

1.60 (100 kHz to 300 kHz)

 

 

 

 

 

 

1.20 (300 kHz to 1 MHz)

 

 

 

 

 

 

1.1 (1 MHz to 2.0 GHz)

 

 

 

 

 

 

1.30 (2.0 to 2.9 GHz)

 

Power Sensor 2

 

Frequency Range: 1 MHz to 2 GHz

HP 8483A

 

 

 

 

Maximum SWR: 1.18 (600 kHz to 2.0 GHz) 75 ￿

 

Power Sensor ,

 

Frequency Range: 300 MHz

HP 8484A

Low Power

 

Amplitude Range:

020 dBm to 070 dBm

 

 

 

 

Maximum SWR:

1.1 (300 MHz)

 

1 Precision

Frequency

Reference only

 

 

Calibrating 1-11

 

 

 

2 HP 8591EM only