Table
Equipment |
|
| Critical Specications for | Recommended | |||
|
|
|
| Equipment Substitution | Model | ||
Digital V oltmeter |
| Input Resistance: | 10 megohms | HP 3456A | |||
|
|
| Accuracy: 610 mV on 100 V range |
| |||
Frequency Counter 1 |
| Frequency: 10 MHz |
| HP 5334A/B | |||
|
|
| Resolution: | 60.002 Hz |
|
| |
|
|
| External Timebase |
|
| ||
Frequency | Standard |
| Frequency: | 10 MHz |
| HP 5061B | |
|
|
| Timebase Accy (Aging): <1 21009/day |
| |||
Measuring | Receiver |
| Compatible | with Power Sensors | HP 8902A | ||
| |||||||
|
|
| dB Relative Mode |
|
|
| |
|
|
| Resolution: | 0.01 dB |
|
| |
|
|
| Reference | Accuracy: 61.2% |
| ||
Microwave Frequency |
| Frequency Range: | 9 MHz to 7 GHz | HP 5343A | |||
| |||||||
Counter |
|
| Timebase Accy (Aging): <5 2 10010/day |
| |||
Power Meter |
| Power Range: Calibrated | in dBm and | HP 436A | |||
|
|
| dB relative to reference | power 070 dBm |
| ||
|
|
| to +44 dBm, sensor dependent |
| |||
Power Sensor |
| Frequency Range: 100 kHz to 1800 MHz | HP 8482A | ||||
| |||||||
|
|
| Maximum SWR: | 1.60 (100 kHz to 300 kHz) |
| ||
|
|
|
|
| 1.20 (300 kHz to 1 MHz) |
| |
|
|
|
|
| 1.1 (1 MHz to 2.0 GHz) |
| |
|
|
|
|
| 1.30 (2.0 to 2.9 GHz) |
| |
Power Sensor 2 |
| Frequency Range: 1 MHz to 2 GHz | HP 8483A | ||||
| |||||||
|
|
| Maximum SWR: 1.18 (600 kHz to 2.0 GHz) 75 |
| |||
Power Sensor , |
| Frequency Range: 300 MHz | HP 8484A | ||||
Low Power |
| Amplitude Range: | 020 dBm to 070 dBm |
| |||
|
|
| Maximum SWR: | 1.1 (300 MHz) |
| ||
1 Precision | Frequency | Reference only |
|
| Calibrating | ||
|
|
|
2 HP 8591EM only