12. Sweep Time Accuracy, HP 8590
11.Change the EMC analyzer's sweep time setting and the synthesizer/function generator's frequency for the next TR entry listed in Table
Table
EMC Analyzer | Synthesizer/Function | Minimum | TR Entry | Maximum |
Sweep Time Setting | Generator Frequency | Reading | (MKR 1) | Reading |
|
|
|
|
|
20 ms | 500.0 Hz | 15.4 ms | 1 | 16.6 ms |
100 ms | 100.0 Hz | 77.0 ms | 2 | 83.0 ms |
1 s | 10.0 Hz | 770.0 ms | 3 | 830.0 ms |
10 s | 1.0 Hz | 7.7 s | 4 | 8.3 s |
|
|
|
|
|