61.Tracking Generator LO Feedthrough Amplitude, HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM Option 010
61.Tracking Generator LO F eedthrough Amplitude , HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM Option 010
This performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
The tracking generator output is connected to the EMC analyzer INPUT 50
and the tracking is adjusted at 300 MHz for a maximum signal level. The tracking generator output is then connected to the input of a microwave spectrum analyzer. The tracking generator is tuned to several dierent frequencies and the LO Feedthrough is measured at the frequency extremes of the LO.
There are no related adjustment procedures for this performance verication test.
Equipment Required
Microwave spectrum analyzer
Cable, Type N, 62 cm (24 in)
Cable, BNC, 23 cm (9 in)
Adapter, Type N (m) to BNC (f)
Figure
LO Feedthrough Amplitude Test Setup,
HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM
Performance Verication Tests