This performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
The frequency readout accuracy of the EMC analyzer is tested with an input signal of known frequency. By using the same frequency standard for the EMC analyzer and the synthesized sweeper, the frequency reference error is eliminated.
Equipment Required
Synthesized sweeper
Adapter, Type N (f) to APC 3.5 (m)
Adapter, APC 3.5 (f) to APC 3.5 (f)
Cable, BNC, 122 cm (48 in)
Cable, APC 3.5 mm (m) 91 cm (36 in)
Figure
Frequency Readout Accuracy Test Setup, HP 8591EM and HP 8594EM
Procedure
This performance verication test consists of two parts:
Part 1: Frequency Readout Accuracy
Part 2: Marker Count Accuracy
Perform \Part 1: Frequency Readout Accuracy" before performing \Part 2: Marker Count Accuracy."