23. Frequency Response, HP 8596EM
Frequency Response , Band 2
26. Press the following EMC analyzer keys:
4FREQUENCY5 More 1 of 2 Band Lock
4FREQUENCY5 6.0 4GHz5
CF STEP AUTO MAN (MAN) 200 4MHz5
4SPAN5 10 4MHz5
4BW5 IF BW AUTO MAN (MAN) 1 4MHz5
AVG BW AUTO MAN (MAN) 10 4kHz5
4MKR !5 MARKER ! HIGH
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
27.Set the synthesized sweeper CW to 6.0 GHz.
28.On the EMC analyzer, press 4AMPLITUDE5, More 1 of 3 , and PRESEL PEAK .
29.Adjust the synthesized sweeper power level for an EMC analyzer
30.Record the negative of the power ratio displayed on the measuring receiver in Table
31.On the synthesized sweeper, press 4CW5 , and 4 * 5 (step up) key, then on the EMC analyzer, press 4FREQUENCY5, and 4 * 5 (step up) key to step through the remaining frequencies listed in Table
32.At each new frequency repeat steps 28 through 30, entering the power sensor Cal Factor into the measuring receiver as indicated in Table
Performance Verication Tests