23. Frequency Response, HP 8596EM

Frequency Response , Band 2

26. Press the following EMC analyzer keys:

4FREQUENCY5 More 1 of 2 Band Lock 6.0-12.8 BAND 2

4FREQUENCY5 6.0 4GHz5

CF STEP AUTO MAN (MAN) 200 4MHz5

4SPAN5 10 4MHz5

4BW5 IF BW AUTO MAN (MAN) 1 4MHz5

AVG BW AUTO MAN (MAN) 10 4kHz5

4MKR !5 MARKER ! HIGH

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

27.Set the synthesized sweeper CW to 6.0 GHz.

28.On the EMC analyzer, press 4AMPLITUDE5, More 1 of 3 , and PRESEL PEAK .

29.Adjust the synthesized sweeper power level for an EMC analyzer MKR-TRK amplitude reading of 93 dB￿V 60.1 dB.

30.Record the negative of the power ratio displayed on the measuring receiver in Table 2-40, column 2.

31.On the synthesized sweeper, press 4CW5 , and 4 * 5 (step up) key, then on the EMC analyzer, press 4FREQUENCY5, and 4 * 5 (step up) key to step through the remaining frequencies listed in Table 2-40.

32.At each new frequency repeat steps 28 through 30, entering the power sensor Cal Factor into the measuring receiver as indicated in Table 2-40.

Performance Veri￿cation Tests 2-145