27. Other Input Related Spurious Responses, HP 8595EM
Figure
Other Input Related Spurious Responses Test Setup, HP 8595EM
Procedure
Band 0
1.Zero and calibrate the measuring receiver and power sensor in log mode (power reads out in dBm), as described in the measuring receiver operation manual. Enter the power sensor 2 GHz Cal Factor into the measuring receiver.
2.Press INSTRUMENT PRESET on the synthesized sweeper and set the controls as follows:
CW :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: :: 2000 MHz
POWER LEVEL : ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: 04 dBm
3.Connect the equipment as shown in Figure
4.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 2.0 4GHz5
4SPAN5 1 4MHz5
4AMPLITUDE5 REF LVL 97 4dBV5
ATTEN AUTO MAN (MAN) 0 4dB5