32. Spurious Response, HP 8595EM
Figure
45.If the distortion products can be seen, proceed as follows:
a.On the EMC analyzer, press 4MKR !5, MARKER ! HIGH , More 1 of 3 ,
More 2 of 3 , and MARKER 1 .
b.Press More 3 of 3 then repeatedly press NEXT PEAK until the active marker is on the highest distortion product.
c.Record the MKR 1 amplitude reading as TR Entry 2 of the performance verication test record. The MKR 1 reading should be less than 054 dBc.