58. Tracking Generator Feedthrough, HP 8591EM Option 010
12.Press 4AUX/USER5, Track Gen , SRC PWR ON OFF (ON), and enter 107 4dBV5.
13.Press 4SGL SWP5, then wait for completion of a new sweep. Press 4DISPLAY5,
DSP LINE ON OFF (ON).
14.Adjust the display line so that it is centered on the average trace noise, ignoring any residual responses. Record the display line amplitude setting in Table
15.Repeat steps 13 and 14 for the remaining Tracking Generator Output Frequencies (EMC analyzer center frequency) listed in Table
16.In Table
Table
Tracking Generator | Noise Level |
Output Frequency | Amplitude |
| (dBV) |
|
|
1MHz
20 MHz
50 MHz
100MHz
250MHz
400MHz
550MHz
700MHz
Tracking Generator | Noise Level |
Output Frequency | Amplitude |
| (dBV) |
|
|
850MHz
1000 MHz
1150 MHz
1300 MHz
1450 MHz
1600 MHz
1750 MHz