36. Gain Compression, HP 8594EM

24.On the EMC analyzer, press 4SGL SWP5, then wait for the completion of a new sweep. Press 4MKR !5 MKR ! HIGH More 1 of 3 More 2 of 3 MARKER 1 .

25.On synthesized sweeper #1, set RF to ON.

26.On the EMC analyzer, press 4SGL SWP5, then wait for the completion of a new sweep. Press 4MKR !5 MKR ! HIGH More 1 of 3 More 2 of 3 MARKER 1 .

27.Read the MKR 1 amplitude and record in the performance veri￿cation test record as TR Entry 2.

2-252 Performance Veri￿cation Tests