62.CISPR Pulse Response, HP 8590
57.Press the following EMC analyzer keys:
4SPAN5 ZERO SPAN
4SWEEP/TRIG5 SWP TIME AUTO MAN (MAN) 2 4SEC5
4DET5 QP ON OFF (ON)
Note that this routine will take approximately 1 minute to execute.
4MKR !5 MARKER ! HIGH
58.Record the marker amplitude reading in:
Table
Table
59.Set the PERIOD to 1 ms on the pulse generator.
60.On the EMC analyzer, press 4MKR !5 MARKER ! HIGH .
61.Record the marker amplitude reading in:
Table
Table
62.Press the following EMC analyzer keys:
4DET5 AVG ON OFF (ON)
4MKR !5 MARKER ! HIGH
63.Record the marker amplitude reading in Table
64.Press the following EMC analyzer keys:
4DET5 QP ON OFF (ON)
4AMPLITUDE5 REF LVL 67 4dBV5
65.Set the PERIOD to 50 ms on the pulse generator.
66.Press 4SGL SWP5 on the EMC analyzer, then press 4MKR !5
MARKER ! HIGH .