59. Tracking Generator Feedthrough, HP 8594EM Option 010
Procedure
1.Connect the Type N cable between the RF OUT 50 and INPUT 50 connectors on the EMC analyzer. See Figure
2.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 300 4MHz5
4SPAN5 0 4Hz5
4BW5 IF BW AUTO MAN (MAN) 30 4kHz5
4MKR5
4AUX/USER5 Track Gen
SRC PWR ON OFF (ON) 102 4dBV5
3.On the EMC analyzer, press TRACKING PEAK . Wait for the PEAKING message to disappear.
4.Connect the CAL OUTPUT to the INPUT 50 .
5.Set the EMC analyzer by pressing the following keys:
4SPAN5 10 4MHz5
4AMPLITUDE5 REF LVL 87 4dBV5
ATTEN AUTO MAN (MAN) 0 4dB5
4MKR !5 MKR ! HIGH
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4SPAN5 100 4kHz5
Wait for the AUTO ZOOM message to disappear, then set the EMC analyzer as follows:
4BW5 AVG BW AUTO MAN (MAN) 30 4Hz5
4MKR5 More 1 of 3 MK TRACK ON OFF (OFF)
6.Press 4SGL SWP5, wait for the completion of a new sweep, then press 4MKR !5
MKR ! HIGH .
Subtract the MKR amplitude reading from 87 dBV, then enter the result in the EMC analyzer as the REF LVL OFFSET. For example, if the marker reads 86.79 dBV, enter +0.21 dB.