4.Frequency Readout and Marker Count Accuracy, HP 8591EM and HP 8594EM
4SPAN5 20 4kHz5
8.Press 4MKR !5 MARKER ! HIGH on the EMC analyzer.
9.Record the MKR frequency reading as TR Entry 4 of the performance verication test record. The reading should be within the limits of 1.49999924 GHz and 1.50000076 GHz.
\Part 1: Frequency Readout Accuracy" is now complete. Continue with \Part 2: Marker Count Accuracy."