53. Tracking Generator Level Flatness, HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM Option 010

20.Locate the most positive Level Flatness readings in Table 2-68 and

Table 2-69 and record these values as TR Entry 1 and TR Entry 2 of the performance veri￿cation test record.

21.Locate the most negative Level Flatness readings in Table 2-68 and Table 2-69 and record this value as TR Entry 3 and TR Entry 4 of the performance veri￿cation test record.

Performance Veri￿cation Tests 2-345