8. Frequency Span Readout Accuracy, HP 8591EM
7. Press More 1 of 3 More 2 of 3 MARKER 1 , then continue pressing
More 3 of 3 NEXT PK RIGHT until the marker 1 is on the
8.Record the MKR 1 frequency reading as TR Entry 1 of the performance verication test record.
The MKR 1 reading should be between 1446 MHz and 1554 MHz.
Figure
10.1MHz to 10 kHz Frequency Span Readout Accuracy Test Setup, HP 8591EM
Part 2: 10.1 MHz to 10 kHz Frequency Span Readout A ccuracy
Perform \Part 1: 1800 MHz Frequency Span Readout Accuracy" before performing this procedure.
1.Connect the equipment as shown in Figure
2.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 70 4MHz5
4SPAN5 10.1 4MHz5