28.Other Input Related Spurious Responses, HP 8596EM
8.Press the following EMC analyzer keys:
4MKR5 MARKER 1 ON OFF (OFF)
4DISPLAY5 HOLD
4AUTO COUPLE5 AUTO ALL
4SPAN5 1 4MHz5
4AMPLITUDE5 REF LVL 97 4dBV5
ATTEN AUTO MAN (MAN) 0 4dB5
4SWEEP5 SWEEP CONT SGL (CONT)
Band 1
9.On the EMC analyzer, press 4FREQUENCY5, 4, 4GHz5.
10.Set the synthesized sweeper CW to 4 GHz.
11.Enter the power sensor 4 GHz CAL Factor into the measuring receiver.
12.Press the following EMC analyzer keys:
4MKR !5 MKR ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK
Wait for the CAL: PEAKING message to disappear, then press:
4MKR5 More 1 of 3 MARKER ALL OFF
13.Repeat steps 5 through 8 for the synthesized sweeper CW frequencies listed in Table
Band 2
14.On the EMC analyzer, press 4FREQUENCY5, 9, 4GHz5.
15.Set the synthesized sweeper CW to 9 GHz.
16.Enter the power sensor 9 GHz CAL Factor into the measuring receiver.
17.Press the following EMC analyzer keys:
4MKR !5 MKR ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK
Performance Verication Tests