32. Spurious Response, HP 8595EM

23.Calculate the desired maximum marker amplitude reading as follows:

a.Add the Frequency Response Error (FRE) to 060 dBc (speci￿cation is 0100 dBc, but reference level will be changed by 40 dB to yield the required dynamic range), then record the Distortion-limited Speci￿cation in Table 2-48.

Distortion-limited Speci￿cation = 060 dBc + FRE

b.Subtract 67 dB￿V (reference level setting) from Noise Level at 5.6 GHz, then record in Table 2-48.

Noise-limited Speci￿cation = Noise Level at 5.6 GHz + 67 dB￿V

c.Record the more positive of the values recorded in a and b above as

TR Entry 2 of the performance veri￿cation test record. For example, if the value in a is 059 dBc and the value in b is 061 dBc, record 059 dBc.

Figure 2-49.

Second Harmonic Distortion Test Setup, >2.9 GHz, HP 8595EM

24.Connect the equipment as shown in Figure 2-49 with the ￿lter in place.

25.Set the synthesized sweeper controls as follows:

CW : : :: :: ::: :: :: :: ::: :: :: :: :: : :: :: :: :: ::: :: :: :: ::: :: :: :: :: :: 2.8 GHz

POWER LEVEL : :: :: :: :: ::: :: :: :: :: ::: :: : : :: ::: :: :: :: ::: :: :: : 0 dBm

Performance Veri￿cation Tests 2-213