46. Residual Responses, HP 8593EM, HP 8595EM, and HP 8596EM
17.Press 4SGL SWP5 and wait for a new sweep to nish. Look for any residual responses at or above the display line.
If a residual is suspected, press 4SGL SWP5 again. A residual response will persist on successive sweeps, but a noise peak will not. Note the frequency and amplitude of any residual responses above the display line in Table
18.Press 4FREQUENCY5, 4 * 5
19.Repeat step 18 until the range from 2.75 GHz to 6.5 GHz has been checked. (This requires 372 additional frequency steps.)
Table
Residual Responses above Display Line Worksheet
| Frequency |
| Amplitude | ||
| (MHz) |
| (dBV) | ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Conrming Residuals
20.Set the EMC analyzer center frequency to a residual frequency recorded in Table
4PRESET5
4AMPLITUDE5 47 4dBV5 ATTEN AUTO MAN (MAN) 0 4dB5 4SPAN5 20 4kHz5
4SGL SWP5
4DISPLAY5 DSP LINE ON OFF (ON) 17 4dBV5
21.Press 4SGL SWP5 and wait for a new sweep to nish. Look for any residual responses at or above the display line.