17.Resolution (IF) Bandwidth Accuracy, HP 8590
This performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
The output of a synthesizer/level generator is connected to the input of the EMC analyzer. Measurements are performed in zero span to reduce the measurement uncertainty.
The frequency of the synthesizer/level generator is set to the center of the
The frequency of the synthesizer is reduced then recorded when the resulting marker amplitude matches the previously set marker reference. The synthesizer frequency is increased so that it is tuned on the opposite point on the skirt of the lter response. The frequency is once again recorded and the dierence between the two frequencies is compared to the specication.
Equipment Required
Synthesizer/level generator
Cable, BNC, 122 cm (48 in)
Adapter, Type N (m) to BNC (f)
Additional Equipment for Option 026
Adapter, APC 3.5 (f) to Type N (f)
Performance Verication Tests