51. Power Sweep Range, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM

SRC PWR ON OFF (ON), and adjust the power level until the marker reads 0 dB 60.1 dB.

8.Record the power level displayed on the measuring receiver as TR Entry 1 of the performance veri￿cation test record.

9. Press PWR SWP ON OFF (ON) to set power sweep on. Wait for completion of a new sweep.

10.Press 4MKR5, MARKER NORMAL . Use the knob to place the marker at the right-most graticule line. Press MARKER 1 .

Performance Veri￿cation Tests 2-333