46. Residual Responses, HP 8593EM, HP 8595EM, and HP 8596EM

1 MHz to 2.75 GHz

4.Connect the 300 MHz CAL OUT to the RF INPUT.

5.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to ￿nish. Press the following keys:

4FREQUENCY5 300 4MHz5 More 1 of 2 Band Lock BAND LOCK ON OFF

(ON)

4SPAN5 10 4MHz5

4MKR !5 MKR ! HIGH

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

4SPAN5 1 4kHz5

Wait for the AUTO ZOOM message to disappear, then press

6. Press the following EMC analyzer keys:

4BW5 300 4Hz5

4SWEEP/TRIG5 1 4SEC5

4AMPLITUDE5 87 4dB￿V5

ATTEN AUTO MAN (MAN) 0 4dB5

7. Press the following EMC analyzer keys:

4SGL SWP5

4MKR !5 MKR ! HIGH More 1 of 3 More 2 of 3 MARKER 1

4SPAN5 10 4MHz5

4SGL SWP5

4MKR !5 MKR ! HIGH

8. Record the marker-1 reading below as the MEAS UNCAL Amplitude Error.

MEAS UNCAL Amplitude Error

 

dB

9.Remove the cable from the EMC analyzer input.

10.Reconnect the termination to the EMC analyzer input as shown in Figure 2-68.

11.Press the following EMC analyzer keys:

4FREQUENCY5 5 4MHz5

4AMPLITUDE5 47 4dB￿V5

2-310 Performance Veri￿cation Tests