46. Residual Responses, HP 8593EM, HP 8595EM, and HP 8596EM
1 MHz to 2.75 GHz
4.Connect the 300 MHz CAL OUT to the RF INPUT.
5.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Press the following keys:
4FREQUENCY5 300 4MHz5 More 1 of 2 Band Lock BAND LOCK ON OFF
(ON)
4SPAN5 10 4MHz5
4MKR !5 MKR ! HIGH
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4SPAN5 1 4kHz5
Wait for the AUTO ZOOM message to disappear, then press
6. Press the following EMC analyzer keys:
4BW5 300 4Hz5
4SWEEP/TRIG5 1 4SEC5
4AMPLITUDE5 87 4dBV5
ATTEN AUTO MAN (MAN) 0 4dB5
7. Press the following EMC analyzer keys:
4SGL SWP5
4MKR !5 MKR ! HIGH More 1 of 3 More 2 of 3 MARKER 1
4SPAN5 10 4MHz5
4SGL SWP5
4MKR !5 MKR ! HIGH
8. Record the
MEAS UNCAL Amplitude Error |
| dB |
9.Remove the cable from the EMC analyzer input.
10.Reconnect the termination to the EMC analyzer input as shown in Figure
11.Press the following EMC analyzer keys:
4FREQUENCY5 5 4MHz5
4AMPLITUDE5 47 4dBV5