33. Spurious Response, HP 8596EM
If necessary, adjust the EMC analyzer center frequency until the two signals are centered on the display.
42. Set the EMC analyzer by pressing the following keys:
4BW5 IF BW AUTO MAN (MAN) 1 4kHz5
AVG BW AUTO MAN (MAN) 100 4Hz5
43. Press the following analyzer keys:
4MKR !5 MARKER ! HIGH More 1 of 3 More 2 of 3 MARKER 1 4DISPLAY5 DSP LINE ON OFF (ON)
Set the display line to a value 54 dB below the current reference level setting.
44.The
Figure