62. CISPR Pulse Response, HP 8590 EM-Series

200 Hz Pulse RF Signal Setup

20.Press 4RECALL5 1 on the pulse generator. Set the pulse generator to the following conditions. (Use the 4CHS5 key to change the sign of the value entered on the pulse generator.)

PER :: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: ::

40 ms

WID : :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: :::

0.1 ms

LOL : :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :

01.7 V

21.Press 4RECALL5 1 on the synthesizer/level generator. Set the synthesizer/level generator amplitude to the required amplitude value for the 200 Hz ￿lter recorded in Table 2-80 by pressing 4AMPLITUDE5, (enter the Required Amplitude for 200 Hz), 40dBm5.

22.Press the following EMC analyzer keys:

4SPAN5 ZERO SPAN

4SWEEP/TRIG5 SWP TIME AUTO MAN (MAN) 2 4SEC5

4DET5 QP ON OFF (ON)

Note that this routine will take approximately 1 minute to execute.

4MKR !5 MARKER ! HIGH

23.Record the marker amplitude reading in:

Table 2-81 as the Measured 25 Hz Amplitude for 200 Hz.

Table 2-82 as the Measured Relative Equivalent Level of Pulse for Band A, 25 Hz Repetition Frequency.

Table 2-83 as the Measured Relative Equivalent Level of Pulse for Band A (QP) 25 Hz Repetition Frequency.

24.Press the following EMC analyzer keys:

4DET5 AVG ON OFF (ON)

4MKR !5 MARKER ! HIGH

25.Record the marker amplitude reading in Table 2-83 as the Measured Relative Equivalent Level of Pulse for Band A (AVG) 25 Hz Repetition Frequency.

26.Press the following EMC analyzer keys:

4DET5 QP ON OFF (ON)

2-390 Performance Veri￿cation Tests