62. CISPR Pulse Response, HP 8590
200 Hz Pulse RF Signal Setup
20.Press 4RECALL5 1 on the pulse generator. Set the pulse generator to the following conditions. (Use the 4CHS5 key to change the sign of the value entered on the pulse generator.)
PER :: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: | 40 ms |
WID : :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: | 0.1 ms |
LOL : :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: : | 01.7 V |
21.Press 4RECALL5 1 on the synthesizer/level generator. Set the synthesizer/level generator amplitude to the required amplitude value for the 200 Hz lter recorded in Table
22.Press the following EMC analyzer keys:
4SPAN5 ZERO SPAN
4SWEEP/TRIG5 SWP TIME AUTO MAN (MAN) 2 4SEC5
4DET5 QP ON OFF (ON)
Note that this routine will take approximately 1 minute to execute.
4MKR !5 MARKER ! HIGH
23.Record the marker amplitude reading in:
Table
Table
Table
24.Press the following EMC analyzer keys:
4DET5 AVG ON OFF (ON)
4MKR !5 MARKER ! HIGH
25.Record the marker amplitude reading in Table
26.Press the following EMC analyzer keys:
4DET5 QP ON OFF (ON)