5. Frequency Readout and Marker Count Accuracy, HP 8593EM, HP 8595EM, and HP 8596EM

12.Record the MKR frequency reading as TR Entry 16 of the performance veri￿cation test record. The reading should be within the limits of 1.49999924 GHz and 1.50000076 GHz.

\Part 1: Frequency Readout Accuracy" is now complete. Continue with \Part 2: Marker Count Accuracy."

Part 2: Marker Count A ccuracy

Perform \Part 1: Frequency Readout Accuracy" before performing this procedure.

1.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to ￿nish.

2.Set the EMC analyzer to measure the marker count accuracy by pressing the following keys:

4FREQUENCY5 1.5 4GHz5

4SPAN5 20 4MHz5

4BW5 IF BW AUTO MAN (MAN) 300 4kHz5

4MKR5 More 1 of 3 MK COUNT ON OFF (ON)

More 2 of 3 CNT RES AUTO MAN (MAN) 100 4Hz5

3.Press 4MKR !5 MARKER ! HIGH , then wait for a count to be taken (it may take several seconds).

4.Record the CNTR frequency reading as TR Entry 17 of the performance veri￿cation test record. The reading should be within the limits shown in Table 2-3.

2-20 Performance Veri￿cation Tests