62. CISPR Pulse Response, HP 8590
FREQUENCY :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: : 50 MHz
AMPLITUDE : :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: : 03 dBm
6. Set the power meter to the following settings:
MODE ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: : dBm
CAL FACTOR : :: :: ::: :: : power sensor Ref Cal Factor for 50 MHz
7.Adjust synthesizer/level generator power level for a 06.99 dBm (60.03) reading on the power meter.
8.Record the synthesizer/level generator amplitude setting in Table
Table
EMI | Reference Amplitude | Amplitude |
| Required | |||
Bandwidth |
| at 50 MHz | Oset |
| Amplitude | ||
|
|
|
|
|
|
|
|
200 Hz |
|
|
| 00.40 |
|
|
|
9 kHz |
|
|
| 0.05 |
|
|
|
120 kHz |
|
|
| 5.42 |
|
|
|
|
|
|
|
|
|
|
|
9.Calculate the Required Amplitude for each EMI bandwidth using the following formula and enter each calculated Required Amplitude values in Table
Reference Amplitude at 50 MHz + Amplitude Oset = Required Amplitude
Note that the reference amplitude is the same for the 200 Hz, 9 kHz, and 120 kHz lters.
10.On the synthesizer/level generator, press 4STORE5 1 to store the previous setting of the synthesizer/level generator in storage register 1.
Performance Verication Tests