48.Fast Time Domain Sweeps, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM Option 101 and Option 301

4MKR !5 MKR ! HIGH

If necessary, press NEXT PEAK or NEXT PK LEFT until the marker is on the left-most complete signal peak. This is the \marked signal."

11.Press More 1 of 3 More 2 of 3 MARKER 1 , MARKER 1 , then press

NEXT PK RIGHT until the marker 1 is on the eighth signal.

12.Record the MKR 1 frequency reading in the performance veri￿cation test record as shown in Table 2-62. The MKR reading should be within the limits shown.

13.Repeat steps 10 through 12 for the remaining sweep time settings listed in Table 2-62.

Table 2-62. Fast Sweep Time Accuracy

EMC Analyzer

Synthesizer/Function

Minimum

TR Entry

Sweep Time

Generator Frequency

Reading

(MKR 1)

 

 

 

 

 

18 ms

556 Hz

14.04

ms

1

10 ms

1 kHz

7.8

ms

2

1.0 ms

10 kHz

780

￿s

3

100 ￿s

100

kHz

78 ￿s

4

20 ￿s

500

kHz

15.6

￿s

5

 

 

 

 

 

 

Performance Veri￿cation Tests 2-321