58.Tracking Generator Feedthrough, HP 8591EM Option 010

58.Tracking Generator F eedthrough,

HP 8591EM Option 010

This performance veri￿cation test must be performed with the EMC analyzer set in the spectrum analyzer mode.

The tracking generator output is connected to the EMC analyzer input and the tracking is adjusted at 300 MHz for a maximum signal level. The tracking generator output is terminated and set for 0 dBm output power (maximum output power). The EMC analyzer input is also terminated. The noise level of the EMC analyzer is then measured at several frequencies.

There are no related adjustments for this performance test.

Equipment Required

50 ￿ Termination (two required) Cable, Type N, 62 cm (24 in) Cable, BNC, 23 cm (9 in)

Cable, Type N (m) to BNC (f)

Figure 2-83. Tracking Generator Feedthrough Test Setup, HP 8591EM

Performance Veri￿cation Tests 2-367