45. Residual Responses, HP 8594EM

45. Residual Responses , HP 8594EM

This performance veri￿cation test must be performed with the EMC analyzer set in the spectrum analyzer mode.

The EMC analyzer input is terminated and the EMC analyzer is swept from 150 kHz to 1 MHz. Then the EMC analyzer is swept in 10 MHz spans throughout the 1 MHz to 2.9 GHz range. Any responses above the speci￿cation are noted.

There are no related adjustment procedures for this performance test.

Equipment

Termination, 50 ￿

Adapter, Type N (m) to APC 3.5 (f)

Figure 2-67. Residual Response Test Setup, HP 8594EM

Procedure

150 kHz to 1 MHz

1.Connect the termination to the EMC analyzer input as shown in Figure 2-67.

2.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to ￿nish. Press the following EMC analyzer keys:

4FREQUENCY5

START FREQ 150 4kHz5

STOP FREQ 1 4MHz5

4AMPLITUDE5 47 4dB￿V5 ATTEN AUTO MAN (MAN) 0 4dB5

Performance Veri￿cation Tests 2-303