16.Absolute Amplitude Calibration and Resolution (IF) Bandwidth Switching Uncertainties,HP 8590
5.Press 4AMPLITUDE5 and use the knob to adjust the reference level until the signal appears one division below the reference level, then press the following keys:
4MKR !5 MARKER ! HIGH
More 1 of 3 More 2 of 3 MARKER 1
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
6.Set the EMC analyzer resolution (IF) bandwidth and span according to Table
Table
Resolution (IF) Bandwidth Switching Uncertainty
EMC Analyzer | MKR 1 TRK Amplitude Reading |
|
|
IF BW Setting SPAN Setting Min. (dB) TR Entry Max. (dB)
3 kHz | 50 kHz | 0 (Ref) | 0 (Ref) | 0 (Ref) |
1 kHz | 50 kHz |
|
|
|
00.5 | 2 | +0.5 | ||
9 kHz | 50 kHz | 00.4 | 3 | +0.4 |
10 kHz | 50 kHz | 00.4 | 4 | +0.4 |
30 kHz | 500 kHz | 00.4 | 5 | +0.4 |
100 kHz | 500 kHz | 00.4 | 6 | +0.4 |
120 kHz | 500 kHz | 00.4 | 7 | +0.4 |
300 kHz | 5 MHz | 00.4 | 8 | +0.4 |
1 MHz | 10 MHz | 00.4 | 9 | +0.4 |
3 MHz | 10 MHz | 00.4 | 10 | +0.4 |
|
|
|
|
|
7.Press 4MKR !5, MARKER ! HIGH , then record the MKR 1 TRK amplitude reading in the performance verication test record as indicated in Table
The amplitude reading should be within the limits shown.
8.Repeat steps 6 through 7 for each of the remaining bandwidth and span settings listed in Table
9.Press the following EMC analyzer keys: