21. Frequency Response, HP 8594EM
21.Enter the power sensor's 50 MHz Cal Factor into the measuring receiver.
22.Adjust the synthesizer/level generator amplitude until the measuring receiver display reads the same value as recorded in step 11. Record the synthesizer/level generator amplitude here and in Table
Synthesizer/Level Generator Amplitude Setting (50 MHz) |
| dBm |
|
23.Replace the power sensor with the 50 termination.
24.Press the following EMC analyzer keys:
4MKR !5 MARKER ! HIGH
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
4MKR5 MARKER 1
Performance Verication Tests