21. Frequency Response, HP 8594EM

21.Enter the power sensor's 50 MHz Cal Factor into the measuring receiver.

22.Adjust the synthesizer/level generator amplitude until the measuring receiver display reads the same value as recorded in step 11. Record the synthesizer/level generator amplitude here and in Table 2-34.

Synthesizer/Level Generator Amplitude Setting (50 MHz)

 

dBm

 

23.Replace the power sensor with the 50 ￿ termination.

24.Press the following EMC analyzer keys:

4MKR !5 MARKER ! HIGH

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

4MKR5 MARKER 1

Performance Veri￿cation Tests 2-127